Excellent reproducibility is obtainable because dial switches with low contact resistance are used. For example, changes in contact resistance is within ± 1.1mΩ at 0.1Ω setting.

YOKOGAWA 279301, 279303 디케이드 저항 박스 이미지

Excellent reproducibility is obtainable because dial switches with low contact resistance are used. For example, changes in contact resistance is within ± 1.1mΩ at 0.1Ω setting.

원산지JAPAN
제조사YOKOGAWA
브랜드YOKOGAWA
무게1kg

Specifications _ 279301

Dielectric Strength

1,000 V AC for one minute between Panel and Circuit
Insulation ResistanceMore than 500MΩ at 500V DC between panel and circuit
Dimension497(W)×116(H)×140(D)mm, excluding protrusions
19.57 x 4.57 x 5.51 in.
WeightApprox. 4.8kg
10.6 lbs
Resistance Range0.1mΩ to 1,111.2100Ω (minimum resistance is 0.1Ω)
Dial Composition0.001Ω x 10 + 0.01Ω x 10 + 0.1Ω x 11 + 1Ω x 10 + 10Ω x 10 + 100Ω x 10
Resolution1mΩ
Accuracy±(0.01Ω + 2mΩ) at temperature of 23 ±2°C, humidity of 45 to 85%, and 0.1W Power Application
Max. Allowable Input Power0.2SW/slep. Within 1W for overall instrument.
Max. Allowable Input Current50mA (100Ω steps), 150mA (10Ω steps), 500mA (10 Ω steps) and 1.5A (0.1Ω steps)
  • High accuracy and stability
  • High reproducibility

Excellent reproducibility is obtainable because dial switches with low contact resistance are used. For example, changes in contact resistance is within ± 1.1mΩ at 0.1Ω setting.

  • 1 mΩ resolution
  • Simple, quick dial operation
  • In-line display for city reading
  • Ideal for calibration of resistance thermometers and brides

Due to its high accuracy and a dial system, various types of resistance thermometers and bridges can be calibrated accurately and promptly.

  • Excellent anti shock and -vibration properties

Specifications _ 279303

Dielectric Strength

2,500 AC for one minute between Panel and Circuit
Insulation ResistanceMore then 1011Ω at 1,000V DC between panel and circuit
Dimension497(W)×116(H)×140(D)mm, excluding protrusions
19.57 x 4.57 x 5.51 in.
WeightApprox. 5.3kg
11.7 lbs
Resistance Range0 to 1,111.110Ω
Dial Composition100Ω x 10 + 1kΩ x 10 + 10kΩ x 10 + 100kΩ x 10 + 1MΩ x 10 + 10MΩ x 10
Resolution100mΩ
Accuracy100Ω, 1kΩ, 10kΩ, 100kΩ Steps - ±(0.05Ω + 0.05mΩ)
1 MΩ and 10MΩ Steps - ±0.2%
(At temperature 23 ±2°C, humidity below 75%, including residual resistance of approx. 0.05Ω).
Max. Allowable Input100 ΩStep100mA
1 ΩStep30mA
10 ΩStep10mA
100 ΩStep3mA (100 to 600 kΩ)
  2000 V (700 kΩ to 1 MΩ)
1 ΩStep2000 V
10 ΩStep2000 V
  • Up to 100MΩ in 100Ω stop
  • Low voltage coefficient

Variation of the resistance value is less than ±0.1% at 1MΩ and 10MΩ steps against 100V application, and less than ±0.04% at 100Ω 1Ω 10Ω, and 100Ω flaps against 10V application.

  • Shock and vibration proof construction
  • Easy-to-read in line indication
  • Best suited for calibration of insulation resistance testers and bridges