The ‘ Isotec h Model 881 Dual Heatpipe Thermocouple Homogeneity Scanner ’ provides a fully automated solution to the problem of measuring thermocouple homogeneity.
| 원산지 | U.K |
|---|---|
| 제조사 | ISOTECH |
| 브랜드 | ISOTECH |
| 무게 | 1kg |
The ‘Isotech Model 881 Dual Heatpipe Thermocouple Homogeneity Scanner’ provides a fully automated solution to the problem of measuring thermocouple homogeneity.
The operation of a thermocouple relies on the Seebeck effect which causes an emf to be generated in any region of a thermo element that is exposed to a temperature gradient. Undesirably, nearly all thermocouples develop non-uniformities (inhomogeneities) in their thermoelements during normal use.
If one is to assess the accuracy of a thermocouple, then the inhomogeneity of the thermocouple is a major concern. Increasingly, users and laboratories want to be able to measure thermocouple inhomogeneity.
- The scanner can be used to determine:
If wire/cable manufacturing processes meet quality standards or tolerances
Whether a thermocouple is damaged or faulty and unfit for use or calibration
- SPECIFICATION
| Operating Temp | 100°C |
| Max Immersion Depth | 1000mm |
| Recommended T/C Diameter | 1.5 to 4mm |
| Scanning Time | < 20 minutes |
| Height | 2685mm |
| Width | 210mm, 795mm incl base |
| Depth | 212mm, 510mm incl base |
| Weight | 25kg (excluding base) |
| Power | 175 watts |