YOKOGAWA 2406E 시리즈 아날로그 절연저항계

Standard Test Conditions - Model 240631

YOKOGAWA 2406E 시리즈 아날로그 절연저항계 이미지

Standard Test Conditions - Model 240631

원산지JAPAN
제조사YOKOGAWA
브랜드YOKOGAWA
무게1kg

모델번호

ModelRatingEffectiveCentralAC VoltageLower limit ofRated Current
Measuring rangeScale ValueMeasuring rangemeasured Ω*1
240631
25 V/5 MΩ
0.001 to 5 MΩ
0.1 MΩ0 to 300 V
0.025 MΩ
1 mA
50 V/10 MΩ0.005 to 10 MΩ0.2 MΩ0.05 MΩ1 mA
240641125 V/20 MΩ0.01 to 20 MΩ0.5 MΩ0.125 MΩ1 mA
240632
125 V/20 MΩ
0.01 to 20 MΩ
0.5 MΩ
0 to 300 V
0.125 MΩ
1 mA
240642250 V/50 MΩ0.01 to 50 MΩ1 MΩ0.25 MΩ1 mA
240633
125 V/20 MΩ
0.01 to 20 MΩ
0.5 MΩ
0 to 600 V
0.125 MΩ
1 mA
250 V/50 MΩ0.01 to 50 MΩ1 MΩ0.25 MΩ1 mA
240643500 V/100 MΩ0.05 to 100 MΩ2 MΩ0.5 MΩ1 mA
240634
250 V/50 MΩ
0.01 to 50 MΩ
1 MΩ
0 to 600 V
0.25 MΩ
1 mA
500 V/100 MΩ0.05 to 100 MΩ2 MΩ0.5 MΩ1 mA
2406441000 V/2000 MΩ1 to 2000 MΩ50 MΩ1 MΩ1 mA*2
240635
250 V/500 MΩ
0.1 to 500 MΩ
10 MΩ0 to 600 V
0.25 MΩ
1 mA*2
500 V/1000 MΩ0.5 to 1000 MΩ20 MΩ0.5 MΩ1 mA*2
2406451000 V/2000 MΩ1 to 2000 MΩ50 MΩ1 MΩ1 mA*2
Non-Back-Light*1 The minimum value at which the rated voltage can be maintained;
*2 0.55 mA in the case of the first effective measuring range

Specifications - Model 240631

Rating

125V/20M~
25V/5M~
50V/10M~
Effective Measuring Range0.001-5M~
0.005-10M~
0.01-20M~
Central Scale Value0.1M~
0.2M~
0.5M~
AC Voltage Measurement0 to 300 V
Lower Limit0.025M~
0.05M~
0.125~
Rated Current1 mA
Approx. Outer Scale Length88 mm
Discharge FunctionThe tester automatically begins discharge when you turn off the MEAS switch. The pointer swings if there is any residual voltage in the circuit under test. You can make sure discharge is complete by checking that the pointer swings back to the infinite (&
Approx. Width120 mm
Approx. Height110 mm
Approx. Depth60 mm
Approx. Weight500 g
BatteriesSix AA (R6P) batteries
Standard AccessoriesBatteries (Qty- 6)
Carrying case (B9075MU) (Qty -1) Earth probe (blake); approx. 1 m long
Earth probes and Line Probes (98007) (Qty - 1)
User's Manual (Qty - 1)- With probe-housing pocket and neck strap

Standard Test Conditions - Model 240631

Ambient Temperature/Humidity Ranges

23 ±5 /45-75% RH
Position of UseHorizontal (5º max. of angle of inclination)
External Magnetic FieldsNone Battery voltage: Within effective voltage range (The pointer must stay within the range indicated by the BAT symbol when the battery check is performed.)
Tolerance Mentioned ConditionsResistance measurement: First effective measuring range = ±5% of reading
Second effective measuring range = ±10% of reading
Infinite and zero scale values: 0.7% max. of scale length
AC voltage: ±10% of maximum scale value
No-load voltage: 130% max. of rated voltage
Rated measuring current: 1 mA (0 to 20%) when in first effective measuring range
Short-circuit current: 12 mA max.

Specifications - Model 240641

Rating

125V/20M~
25V/5M~
50V/10M~
Effective Measuring Range0.001-5M~
0.005-10M~
0.01-20M~
Central Scale Value0.1M~
0.2M~
0.5M~
AC Voltage Measurement0 to 300 V
Lower Limit0.025M~
0.05M~
0.125~
Rated Current1 mA
Approx. Outer Scale Length88 mm
Discharge FunctionThe tester automatically begins discharge when you turn off the MEAS switch. The pointer swings if there is any residual voltage in the circuit under test. You can make sure discharge is complete by checking that the pointer swings back to the infinite (&
Approx. Width120 mm
Approx. Height110 mm
Approx. Depth60 mm
Approx. Weight500 g
BatteriesSix AA (R6P) batteries
Standard AccessoriesBatteries (Qty- 6)
Carrying case (B9075MU) (Qty -1) Earth probe (blake); approx. 1 m long
Earth probes and Line Probes (98007) (Qty - 1)
User's Manual (Qty - 1)- With probe-housing pocket and neck strap

Standard Test Conditions - Model 240641

Ambient Temperature/Humidity Ranges

23 ±5 /45-75% RH
Position of UseHorizontal (5º max. of angle of inclination)
External Magnetic FieldsNone Battery voltage: Within effective voltage range (The pointer must stay within the range indicated by the BAT symbol when the battery check is performed.)
Tolerance Mentioned ConditionsResistance measurement: First effective measuring range = ±5% of reading
Second effective measuring range = ±10% of reading
Infinite and zero scale values: 0.7% max. of scale length
AC voltage: ±10% of maximum scale value
No-load voltage: 130% max. of rated voltage
Rated measuring current: 1 mA (0 to 20%) when in first effective measuring range
Short-circuit current: 12 mA max.

Specifications - Model 240632

Rating

125V/20M~
Effective Measuring Range0.01-20M~
Central Scale Value0.5M~
AC Voltage Measurement0 to 300 V
Lower Limit0.125~
Rated Current1 mA
Approx. Outer Scale Length88 mm
Discharge FunctionThe tester automatically begins discharge when you turn off the MEAS switch. The pointer swings if there is any residual voltage in the circuit under test. You can make sure discharge is complete by checking that the pointer swings back to the infinite (&
Approx. Width120 mm
Approx. Height110 mm
Approx. Depth60 mm
Approx. Weight500 g
BatteriesSix AA (R6P) batteries
Standard AccessoriesBatteries (Qty- 6)
Carrying case (B9075MU) (Qty -1) Earth probe (blake); approx. 1 m long
Earth probes and Line Probes (98007) (Qty - 1)
User's Manual (Qty - 1)- With probe-housing pocket and neck strap

Standard Test Condtions - Model 240632

Ambient Temperature/Humidity Ranges

23 ±5 /45-75% RH
Position of UseHorizontal (5º max. of angle of inclination)
External Magnetic FieldsNone Battery voltage: Within effective voltage range (The pointer must stay within the range indicated by the BAT symbol when the battery check is performed.)
Tolerance Mentioned ConditionsResistance measurement: First effective measuring range = ±5% of reading
Second effective measuring range = ±10% of reading
Infinite and zero scale values: 0.7% max. of scale length
AC voltage: ±10% of maximum scale value
No-load voltage: 130% max. of rated voltage
Rated measuring current: 1 mA (0 to 20%) when in first effective measuring range
Short-circuit current: 12 mA max.

Specifications - Model 240642

Rating

250V/50M~
Effective Measuring Range0.01-50M~
Central Scale Value1M~
AC Voltage Measurement0 to 300 V
Lower Limit0.25~
Rated Current1 mA
Approx. Outer Scale Length88 mm
Discharge FunctionThe tester automatically begins discharge when you turn off the MEAS switch. The pointer swings if there is any residual voltage in the circuit under test. You can make sure discharge is complete by checking that the pointer swings back to the infinite (&
Approx. Width120 mm
Approx. Height110 mm
Approx. Depth60 mm
Approx. Weight500 g
BatteriesSix AA (R6P) batteries
Standard AccessoriesBatteries (Qty- 6)
Carrying case (B9075MU) (Qty -1) Earth probe (blake); approx. 1 m long
Earth probes and Line Probes (98007) (Qty - 1)
User's Manual (Qty - 1)- With probe-housing pocket and neck strapㅗ

Standard Test Conditions - Model 240642

Ambient Temperature/Humidity Ranges

23 ±5 /45-75% RH
Position of UseHorizontal (5º max. of angle of inclination)
External Magnetic FieldsNone Battery voltage: Within effective voltage range (The pointer must stay within the range indicated by the BAT symbol when the battery check is performed.)
Tolerance Mentioned ConditionsResistance measurement: First effective measuring range = ±5% of reading
Second effective measuring range = ±10% of reading
Infinite and zero scale values: 0.7% max. of scale length
AC voltage: ±10% of maximum scale value
No-load voltage: 130% max. of rated voltage
Rated measuring current: 1 mA (0 to 20%) when in first effective measuring range
Short-circuit current: 12 mA max.

Specifications - Model 240633

Rating

125V/20M~
250V/50M~
500V/100M~
Effective Measuring Range0.01-20M~
0.01-50M~
0.05-100M~
Central Scale Value0.5M~
1M~
2M~
AC Voltage Measurement0 to 600 V
Lower Limit0.125M~
0.25M~
0.5~
Rated Current1 mA
Approx. Outer Scale Length88 mm
Discharge FunctionThe tester automatically begins discharge when you turn off the MEAS switch. The pointer swings if there is any residual voltage in the circuit under test. You can make sure discharge is complete by checking that the pointer swings back to the infinite (&
Approx. Width120 mm
Approx. Height110 mm
Approx. Depth60 mm
Approx. Weight500 g
BatteriesSix AA (R6P) batteries
Standard AccessoriesBatteries (Qty- 6)
Carrying case (B9075MU) (Qty -1) Earth probe (blake); approx. 1 m long
Earth probes and Line Probes (98007) (Qty - 1)
User's Manual (Qty - 1)- With probe-housing pocket and neck strap

Standard Test Conditions - Model 240633

Ambient Temperature/Humidity Ranges

23 ±5 /45-75% RH
Position of UseHorizontal (5º max. of angle of inclination)
External Magnetic FieldsNone Battery voltage: Within effective voltage range (The pointer must stay within the range indicated by the BAT symbol when the battery check is performed.)
Tolerance Mentioned ConditionsResistance measurement: First effective measuring range = ±5% of reading
Second effective measuring range = ±10% of reading
Infinite and zero scale values: 0.7% max. of scale length
AC voltage: ±10% of maximum scale value
No-load voltage: 130% max. of rated voltage
Rated measuring current: 1 mA (0 to 20%) when in first effective measuring range
Short-circuit current: 12 mA max.

Specifications - Model 240343

Rating

125V/20M~
250V/50M~
500V/100M~
Effective Measuring Range0.01-20M~
0.01-50M~
0.05-100M~
Central Scale Value0.5M~
1M~
2M~
AC Voltage Measurement0 to 600 V
Lower Limit0.125M~
0.25M~
0.5~
Rated Current1 mA
Approx. Outer Scale Length88 mm
Discharge FunctionThe tester automatically begins discharge when you turn off the MEAS switch. The pointer swings if there is any residual voltage in the circuit under test. You can make sure discharge is complete by checking that the pointer swings back to the infinite (&
Approx. Width120 mm
Approx. Height110 mm
Approx. Depth60 mm
Approx. Weight500 g
BatteriesSix AA (R6P) batteries
Standard AccessoriesBatteries (Qty- 6)
Carrying case (B9075MU) (Qty -1) Earth probe (blake); approx. 1 m long
Earth probes and Line Probes (98007) (Qty - 1)
User's Manual (Qty - 1)- With probe-housing pocket and neck strap

Standard Test Conditions - Model 240643

Ambient Temperature/Humidity Ranges

23 ±5 /45-75% RH
Position of UseHorizontal (5º max. of angle of inclination)
External Magnetic FieldsNone Battery voltage: Within effective voltage range (The pointer must stay within the range indicated by the BAT symbol when the battery check is performed.)
Tolerance Mentioned ConditionsResistance measurement: First effective measuring range = ±5% of reading
Second effective measuring range = ±10% of reading
Infinite and zero scale values: 0.7% max. of scale length
AC voltage: ±10% of maximum scale value
No-load voltage: 130% max. of rated voltage
Rated measuring current: 1 mA (0 to 20%) when in first effective measuring range
Short-circuit current: 12 mA max.

Specifications - Model 240634

Rating

1000V/2000M~
250V/50M~
500V/100M~
Effective Measuring Range0.01-50M~
0.05-100M~
1-2000M~
Central Scale Value1M~
2M~
50M~
AC Voltage Measurement0 to 600 V
Lower Limit0.25M~
0.5M~
1~
Rated Current1 mA
Approx. Outer Scale Length88 mm
Discharge FunctionThe tester automatically begins discharge when you turn off the MEAS switch. The pointer swings if there is any residual voltage in the circuit under test. You can make sure discharge is complete by checking that the pointer swings back to the infinite (&
Approx. Width120 mm
Approx. Height110 mm
Approx. Depth60 mm
Approx. Weight500 g
BatteriesSix AA (R6P) batteries
Standard AccessoriesBatteries (Qty- 6)
Carrying case (B9075MU) (Qty -1) Earth probe (blake); approx. 1 m long
Earth probes and Line Probes (98007) (Qty - 1)
User's Manual (Qty - 1)- With probe-housing pocket and neck strap

Standard Test Condtions - Model 240634

Ambient Temperature/Humidity Ranges

23 ±5 /45-75% RH
Position of UseHorizontal (5º max. of angle of inclination)
External Magnetic FieldsNone Battery voltage: Within effective voltage range (The pointer must stay within the range indicated by the BAT symbol when the battery check is performed.)
Tolerance Mentioned ConditionsResistance measurement: First effective measuring range = ±5% of reading
Second effective measuring range = ±10% of reading
Infinite and zero scale values: 0.7% max. of scale length
AC voltage: ±10% of maximum scale value
No-load voltage: 130% max. of rated voltage
Rated measuring current: 1 mA (0 to 20%) when in first effective measuring range
Short-circuit current: 12 mA max.

Specifications - 240644

Rating

1000V/2000M~
250V/50M~
500V/100M~
Effective Measuring Range0.01-50M~
0.05-100M~
1-2000M~
Central Scale Value1M~
2M~
50M~
AC Voltage Measurement0 to 600 V
Lower Limit0.25M~
0.5M~
1~
Rated Current1 mA
Approx. Outer Scale Length88 mm
Discharge FunctionThe tester automatically begins discharge when you turn off the MEAS switch. The pointer swings if there is any residual voltage in the circuit under test. You can make sure discharge is complete by checking that the pointer swings back to the infinite (&
Approx. Width120 mm
Approx. Height110 mm
Approx. Depth60 mm
Approx. Weight500 g
BatteriesSix AA (R6P) batteries
Standard AccessoriesBatteries (Qty- 6)
Carrying case (B9075MU) (Qty -1) Earth probe (blake); approx. 1 m long
Earth probes and Line Probes (98007) (Qty - 1)
User's Manual (Qty - 1)- With probe-housing pocket and neck strap

Standard Test Conditions - Model 240644

Ambient Temperature/Humidity Ranges

23 ±5 /45-75% RH
Position of UseHorizontal (5º max. of angle of inclination)
External Magnetic FieldsNone Battery voltage: Within effective voltage range (The pointer must stay within the range indicated by the BAT symbol when the battery check is performed.)
Tolerance Mentioned ConditionsResistance measurement: First effective measuring range = ±5% of reading
Second effective measuring range = ±10% of reading
Infinite and zero scale values: 0.7% max. of scale length
AC voltage: ±10% of maximum scale value
No-load voltage: 130% max. of rated voltage
Rated measuring current: 1 mA (0 to 20%) when in first effective measuring range
Short-circuit current: 12 mA max.

Specifications - Model 240635

Rating

1000V/2000M~
250V/500M~
500V/1000M~
Effective Measuring Range0.1-500M~
0.5-1000M~
1-2000M~
Central Scale Value10M~
20M~
50M~
AC Voltage Measurement0 to 600 V
Lower Limit0.25M~
0.5M~
1~
Rated Current1 mA
Approx. Outer Scale Length88 mm
Discharge FunctionThe tester automatically begins discharge when you turn off the MEAS switch. The pointer swings if there is any residual voltage in the circuit under test. You can make sure discharge is complete by checking that the pointer swings back to the infinite (&
Approx. Width120 mm
Approx. Height110 mm
Approx. Depth60 mm
Approx. Weight500 g
BatteriesSix AA (R6P) batteries
Standard AccessoriesBatteries (Qty- 6)
Carrying case (B9075MU) (Qty -1) Earth probe (blake); approx. 1 m long
Earth probes and Line Probes (98007) (Qty - 1)
User's Manual (Qty - 1)- With probe-housing pocket and neck strap

Standard Test Conditions - Model 240635

Ambient Temperature/Humidity Ranges

23 ±5 /45-75% RH
Position of UseHorizontal (5º max. of angle of inclination)
External Magnetic FieldsNone Battery voltage: Within effective voltage range (The pointer must stay within the range indicated by the BAT symbol when the battery check is performed.)
Tolerance Mentioned ConditionsResistance measurement: First effective measuring range = ±5% of reading
Second effective measuring range = ±10% of reading
Infinite and zero scale values: 0.7% max. of scale length
AC voltage: ±10% of maximum scale value
No-load voltage: 130% max. of rated voltage
Rated measuring current: 1 mA (0 to 20%) when in first effective measuring range
Short-circuit current: 12 mA max.

Specifications - Model 240645

Rating

1000V/2000M~
250V/500M~
500V/1000M~
Effective Measuring Range0.1-500M~
0.5-1000M~
1-2000M~
Central Scale Value10M~
20M~
50M~
AC Voltage Measurement0 to 600 V
Lower Limit0.25M~
0.5M~
1~
Rated Current1 mA
Approx. Outer Scale Length88 mm
Discharge FunctionThe tester automatically begins discharge when you turn off the MEAS switch. The pointer swings if there is any residual voltage in the circuit under test. You can make sure discharge is complete by checking that the pointer swings back to the infinite (&
Approx. Width120 mm
Approx. Height110 mm
Approx. Depth60 mm
Approx. Weight500 g
BatteriesSix AA (R6P) batteries
Standard AccessoriesBatteries (Qty- 6)
Carrying case (B9075MU) (Qty -1) Earth probe (blake); approx. 1 m long
Earth probes and Line Probes (98007) (Qty - 1)
User's Manual (Qty - 1)- With probe-housing pocket and neck strap

Standard Test Conditions - Model 240645

Ambient Temperature/Humidity Ranges

23 ±5 /45-75% RH
Position of UseHorizontal (5º max. of angle of inclination)
External Magnetic FieldsNone Battery voltage: Within effective voltage range (The pointer must stay within the range indicated by the BAT symbol when the battery check is performed.)
Tolerance Mentioned ConditionsResistance measurement: First effective measuring range = ±5% of reading
Second effective measuring range = ±10% of reading Infinite and zero scale values: 0.7% max. of scale length
AC voltage: ±10% of maximum scale value
No-load voltage: 130% max. of rated voltage
Rated measuring current: 1 mA (0 to 20%) when in first effective measuring range
Short-circuit current: 12 mA max.